Super Series Glove Box

  • The Super Glove Box is commonly used in Research & Development Applications and Prototyping in Universities and Commercial Laboratories.
 
  • Glove box function:Self diagnosis, power-off self start, pressure control and adaptation, automatic control, cycle control and password protection, etc.
 
  • Single workstation system: 
  1. Equipped with an integrated gas purification system (single column), PLC controller (with touch screen HMI), electronic casing, front chambers of various sizes, replaceable mirrors and vacuum pump;
  2. The purity can reach H2O<1ppm, O2<1ppm;
  3. Equipped with a large capacity purification system that can adsorb 60 liters of oxygen and absorb 2000 grams of water, thereby extending the time between regeneration cycles.
Availability:
Quantity:
  • Mikrouna

Product Features


Optional IoT smart system enables mobile control and centralized management of the glove box, ensuring safety and peace of mind. It offers intelligent functions including real-time synchronization of device status, equipment data report recording, historical data querying, proactive alarm notifications, remote device debugging, online fault diagnosis and repair requests, as well as online customer support—all accessible through the device.


  • Control System: Siemens PLC


  • Purifier: Single-column


  • Water content: <1 ppm


  • Oxygen content: <1 ppm


  • Leak rate of the glove box can be manually monitored.


Product Specifications


  • Glove Box Chamber


  • Chamber

  1. Material: 304 Stainless Steel, 3 mm thick

  2. Exterior Surface: 304 Stainless Steel

  3. Dimensions: 1220 mm (Length) x 750 mm (Depth) x 900 mm (Height)



  • Front Window

   Tilted Viewing Window: Transparent tempered safety glass, 8 mm thick


  • Glove Port

   Material: POM or Aluminum Alloy: O-ring Seal


  • Gloves

   Butyl Rubber: 0.4 mm thick, 7" or 8" in diameter


  • Filter


  Specifications: 0.3 μm (filtration rating), with one gas inlet and one gas outlet.


  • Shelf

  Material: Stainless Steel; 2 built-in shelves, height-adjustable


  • Chamber Lighting

  Configuration: LED lamps housed within a light cover.


  • Connections

  Specifications: DN40 KF flange, several spare ports, one power supply connection


Purification System


  • Purification Column

  1. Function: Gas-tight sealing and removal of water & oxygen.

  2. Container Material: 304 Stainless Steel

  3. Purification Materials: 5 kg copper catalyst, 5 kg molecular sieve

  4. Purification Capacity: Oxygen removal 60 L, water removal 2 kg.

  5. Specifications: Water and oxygen content < 1 ppm.


  • Circulation System

   Working Gases: Nitrogen, Argon, Helium


  • Regeneration

  1. Operation: Regeneration process is automatically controlled by the PLC.

  2. Regeneration Gas: Working gas mixed with 5–10% hydrogen


  • Vacuum Pump


  1. Specifications (Flow Rate): 8 m³/h, 12 m³/h (Optional)

  2. Rotary Vane Pump: Equipped with an oil mist filter and pneumatic vibration control




  • Valve

  1. Main Valve: DN40 KF electro-pneumatic angle valve.

  2. Control Valve: Integrated solenoid valve.


Antechamber


  • Large Antechamber

  1. Antechamber Body


  • Dimensions: Diameter 360 mm, length 600 mm (Diameter 400 mm optional).

  • Material: 304 Stainless Steel

  • Surface: Interior features an oil-coated brushed finish; exterior is available in either painted or mirror-finished options.

   b.Accessories

  • Tray: Sliding Tray

  • Chamber Door: Double doors made of anodized aluminum alloy, 10 mm thick, vertically operated with a lifting mechanism.

  • Pressure Gauge: Analog display

   c. Control 

  • Operation: Automatic Control


  • Small Antechamber

  1. Antechamber Body

  • Dimensions: Diameter 150 mm, length 300 mm (Diameter 100 mm optional).

  • Material: 304 Stainless Steel

  • Surface: Interior features an oil-coated brushed finish; exterior is available in either painted or mirror-finished options.

   b.Accessories

  •  Antechamber Door: Double door

  • Pressure Gauge: Analog display

   c. Control

  • Operation: Manual Operation


Control System


  • Display: Siemens PLC Touch Screen


  • Functions: Includes self-diagnostics, automatic restart after power failure, pressure control and adaptive functions, automatic control, circulation control, and password protection.


  • Pressure Control: Manages the pressure in both the main chamber and the antechamber. The working pressure of the main chamber can be freely set within a range of ±15 mbar; the system will activate automatic protection if the pressure deviates beyond ±16 mbar.


  • Footswitch: Chamber pressure can be adjusted via the footswitch


  • Data Logger: Automatically records system data.


Accessories

  • Vacuum Pump

    Specifications: 8 m³/h, 12 m³/h, 16 m³/h 


  • Oxygen Analyzer

    Measuring Range: 0–1000 ppm


  • Dew Point Analyzer

   Measuring Range: 0–500 ppm





Related News
16 / 02 / 2026
A Step-by-Step Guide To Glove Box Leak Testing And Pressure Management

A glove box leak rarely announces itself with noise, visible cracks, or dramatic failure.

Read More
23 / 02 / 2026
How To Determine If The Glove Box Needs Maintenance?

Most systems do not fail overnight. A glove box rarely “breaks” in a dramatic way. Instead, it drifts slowly out of optimal condition until research data becomes inconsistent, gas consumption increases, alarms appear more frequently, or operators notice unusual sounds during operation.

Read More
25 / 02 / 2026
The Science of Moisture Control: How Purified Glove Boxes Protect Sensitive Samples

“If your glove box reads low moisture but your samples still degrade, the problem usually isn’t the display—it’s the source of moisture.”

Read More
19 / 02 / 2026
Inert Atmosphere Vs. Vacuum Glove Boxes: Which One Does Your Lab Need?

Many laboratories invest in a glove box believing that adding vacuum capability automatically guarantees a cleaner or more advanced environment.

Read More
18 / 02 / 2026
How To Choose The Right Glove Box for Battery Research: A Complete Guide?

Yesterday your electrode coating looked perfect. Today the electrochemical performance shifts, impedance increases, and the results are inconsistent.

Read More
20 / 01 / 2026
Which Is Better: A Square Or A Circular Transfer Chamber for A Glove Box?

When designing a glove box for specialized tasks such as handling hazardous materials, contaminant-sensitive experiments, or industrial processes, one of the most important features to consider is the transfer chamber.