A The upper alarm system set point of the
Mikrouna glove box is+16mbar; Set the alarm system to -16mbar. When the pressure in the box is higher than+16mbar, the air supply valve will be automatically closed, and the pressure relief exhaust valve (cleaning exhaust valve) will be opened to exhaust. When the pressure is lower than 15mbar, the pressure relief exhaust valve will be closed to automatically restore normal operation; When the pressure of the box is lower than the lower alarm set point -16mbar, the
vacuum pump and exhaust valve are closed, and all system functions are turned off. The pressure relief exhaust valve (cleaning exhaust valve) automatically opens, and when it is higher than -16mbar, the pressure relief valve automatically closes. If there are frequent occurrences of abnormal high or low pressure, the cause must be identified and repaired before the system can be restarted.