  • Q What should be noted when transferring items from the glove box to the outside using the glove box transition compartment?

    A When transferring items from the glove box to the outside using the glove box transition compartment, first check if the outer door is closed properly, reset the pressure gauge pointer to zero, and then confirm that the "gas replenishment" button is closed and that the atmosphere in the transition compartment is consistent with that in the glove box. Only then can items be placed from the glove box into the transition compartment.
  • Q What are the storage operation steps for the glove box large transition compartment in Mikrouna?

    A
    The storage operation steps for the large transition compartment of the Mikrouna glove box are as follows:
    1. Hold the transition compartment door handle tightly with both hands and rotate counterclockwise until the door and body are completely detached.
    2. Hand hold the crossbeam of the large transition compartment and slowly lift the door upwards.
    3. Extract the tray containing the items from the large transition compartment, place the required items onto the tray, and push the tray back into the large transition compartment.
    4. Hand hold the crossbeam of the large transition compartment and press down on the door. When the crossbeam falls on the horizontal axis, the door and the transition compartment are completely aligned. Hold the handle of the transition compartment tightly and rotate it clockwise until the door and body of the transition compartment are completely closed.
    5. Click on the "transition compartment" button on the touchscreen homepage and press the "Vacuum Operation" button.
    6. When the pressure gauge value of the large transition compartment is -0.1Mpa, click the "Add Gas" button on the screen.
    7. Observe the pressure gauge until it returns to zero, then close the air supply button. This is a complete manual operation for the large transition compartment. When inserting general items, repeat this operation three times before opening the inner chamber door, removing the items from the inner door, and then closing the inner chamber door clockwise.
  • Q What should you do if the bottle cap is not properly closed when placing items in the glove box transition compartment?

    A When placing items in the transition compartment of the glove box, if the bottle cap is not properly closed, the transition operation should be stopped immediately. Remove the bottle, replace the cap, and place it in the transition compartment for vacuum treatment again. If the bottle cap of the item placed in the glove box transition compartment is not properly closed, the liquid or gas inside the bottle during vacuuming may cause leakage, pollute the environment inside the transition compartment, or damage other items.
  • Q What is the automatic operation of the transition compartment in Mikrouna's glove box like?

    A The automatic operation of the glove box transition compartment generally refers to the automatic operation of the large transition compartment, while the small transition compartment can only be manually operated. When operating the automatic transition compartment of the Mikrouna glove box, first check if the setting value is correct, then click the automatic operation button of the compartment, and the system will execute according to the set time (the pumping and inflation time can be set, and the number of repetitions can also be set). The default setting for the automatic operation of the large transition compartment in the Mikrouna glove box is 150 seconds for pumping and 75 seconds for inflation, repeated 3 times.
  • Q How can items that cannot withstand vacuum be transferred to the inside of the glove box?

    A If there are special items (unable to withstand a vacuum degree of -0.1Mpa), the vacuuming time of the glove box transition compartment can be shortened to keep the vacuum degree within its acceptable range, but the number of times the transition compartment is filled must be increased.
  • Q Multiple glove boxes are connected, and one of them is faulty. Can the other glove boxes still be used?

    A Multiple glove boxes are connected, one of which is faulty, while the other glove boxes are under program security protection and cannot be used. Two or more glove boxes connected by a T-shaped transition compartment. If one of the glove boxes experiences a power outage or abnormal box pressure, and the T-shaped transition compartment or glove box connected to it is in program safety protection and cannot be used, this type of connected glove box belongs to an integrated device and must ensure that all devices are powered on and operate normally.
  • Q Can the fan of the glove box in Mikrouna be variable frequency?

    A The fan of the Mikrouna glove box can be controlled by variable frequency, and the circulating fan can be set to variable frequency or fixed frequency. In the variable frequency state, the operating frequency of the fan is switched according to the water oxygen value. The smaller the water oxygen value, the lower the frequency, and the slower the fan speed, achieving the goal of energy saving. At the same time, the E00 energy-saving flag will appear next to the fan on the main screen, and the system defaults to variable frequency control. In a fixed frequency state, the fan operates at a fixed high frequency, resulting in fast purification speed.
  • Q Can the vacuum pump of the Mikrouna glove box be timed?

    A The vacuum pump of the Mikrouna glove box can be set to timed or normally open. In timed mode, when the system does not require the vacuum pump to work, it will be turned off after a 10 minute delay and automatically turned on when needed. The system defaults to timed vacuum pump operation. Normally open: As long as the vacuum pump button is turned on, the vacuum pump will remain in the on state.