A Some functions of the
glove box require maintenance to ensure the accuracy and consistency of experiments, including:
1: Regularly check the integrity and sealing of
gloves in the glove box, as gloves are the key part that isolates the glove box from the outside world.
2: It is necessary to regularly replace the gas purification materials of the glove box purification column, such as molecular sieves and activated carbon, to maintain the stability of the gas environment inside the box.
3: Check if the circulating fan and gas pipeline are operating properly to ensure there are no leaks.
4: The sensors inside the box should be calibrated regularly to ensure accurate measurement of water oxygen content.
5: For experiments that generate a large amount of organic vapor, it is recommended to shut down the cycle and analyzer for operation, and then clean the glove box and restart the cycle system after the experiment is completed.
6: Ensure that all operators have received professional training on the correct operating methods and safety regulations for glove boxes.