+86 13600040923        sales.lib@mikrouna.com
You are here: Home / Service / FAQ

FAQ

  • Q What are the components of the purification unit in Mikrouna glove box?

    A The purification unit of Mikrouna glove box mainly consists of purification column, circulation system, regeneration system, vacuum pump, valves and other components. The purification column is equipped with materials such as copper catalyst and molecular sieve, which are used to remove moisture and oxygen from the glove box. The circulation system transports the gas inside the glove box to the purification column for purification through fans and pipelines. The regeneration system recovers its adsorption capacity by heating and circulating specific gases (such as hydrogen) after the purification material is saturated with adsorption. A vacuum pump is used to provide a negative pressure environment for the glove box when needed. The design of the entire purification unit aims to ensure the purity of the internal environment of the glove box and meet the needs of various experiments.
  • Q What is the control system of Mikrouna glove box?

    A The control system of Mikrouna glove box adopts advanced PLC (Programmable Logic Controller) system, which can achieve precise control and monitoring of the internal environment of the glove box. The PLC system, in conjunction with the touch screen interface, allows users to easily set and adjust the working parameters of the glove box, such as gas flow rate, pressure, temperature, etc. In addition, the control system also has self diagnostic function, power-off self starting feature, as well as pressure control and adaptive function, ensuring the stable operation and convenient operation of the glove box.
  • Q What should I do if the purification column temperature is very high during glove box regeneration?

    A During the regeneration process of the glove box, the purification column may generate high temperatures. It is not recommended to operate under high temperature conditions. It is necessary to wait for the purification column to cool naturally for at least 10 hours before operating the glove box to avoid harm to operators or damage to equipment caused by high temperature.
  • Q What is the technical strength of Mikrouna glove box?

    A Mikrouna Glove Box Company and its subsidiaries have over 700 employees, including more than 190 engineering and technical personnel and over 400 technical workers. The core management and technical personnel of the company mainly graduated from well-known domestic universities such as Shanghai Jiao Tong University, Xi'an Jiaotong University, Jilin University, etc., and have more than 170 patented technologies. The company is a national high-tech enterprise and has passed the national level "Specialized, Refined, Unique and New Little Giant" enterprise certification.
  • Q What should I do if there is a system software update or compatibility issue with the glove box control screen?

    A
    When there is a system software update or compatibility issue on the control screen of the glove box, the following measures can be taken:
    1. Check software version: Confirm the current system software version to see if any new updates or patches are available.
    2. Download updates: Download the latest software updates from the manufacturer's official website and follow the instructions for installation.
    3. Contact technical support: If unsure how to operate or encounter problems, contact the technical support team of the glove box manufacturer for assistance.
  • Q What should I do if the pressure inside the glove box fluctuates too much and becomes unstable?

    A
    Excessive pressure fluctuations inside the glove box may be caused by the following reasons:
    1. Check sealing: Ensure that the glove box has good sealing performance and no leakage points.
    2. Check valves: Check if the intake and exhaust valves of the glove box are working properly and if there are any malfunctions or blockages.
    3. Check the pressure regulation system: If the glove box is equipped with a pressure regulation system, check if it is working properly and can effectively maintain stable pressure inside the box.
    4. Check the fan: If the circulating fan inside the glove box is working abnormally, it may cause pressure fluctuations, and it is necessary to check whether the fan needs maintenance or replacement.
    5. If the problem still cannot be solved, it is recommended to contact the manufacturer or professional technicians of the glove box for inspection and repair.
  • Q What should I do if the glove box touch screen operating system responds slowly or malfunctions?

    A
    When the touch screen operating system of the glove box shows slow response or malfunctions, the following measures can be taken for troubleshooting and resolution:
    1. Check power supply: Ensure stable power supply and meet the needs of the touch screen.
    2. Cleaning the surface of the touch screen: Regularly clean the surface of the touch screen to remove dust, dirt, etc., in order to improve response speed.
    3. Update driver: Attempt to update the touch screen driver to address potential software compatibility issues.
    4. Check hardware connection: Ensure that the connection between the touch screen and the computer is normal, without poor contact or short circuits.
    If none of the above methods can solve the problem, it is recommended to contact the touch screen manufacturer or professional technicians for further diagnosis and repair.
  • Q What are the design features of the glove box adsorber?

    A The adsorber design of the glove box takes into account the requirements of efficient adsorption and easy replacement. It is usually located at the front end of the circulation system, allowing the circulating gas to first pass through the adsorber, effectively removing organic solvents. In addition, the design of the adsorber allows for rapid replacement of adsorbent materials without disrupting the high-purity atmosphere inside the glove box, which is particularly important for experimental environments that require frequent replacement of adsorbent materials.