A The service life of the adsorbent material in the
glove box purification column is usually determined based on the degree of decline in its adsorption performance. The effectiveness can be evaluated by regularly monitoring the oxygen and moisture levels inside the glove box. If the level of pollutants inside is found to be abnormally high and the glove box cannot recover its adsorption performance after regeneration, even if the adsorption material has not reached the manufacturer's recommended replacement time, replacement should be considered. In addition, changes in the physical state of the adsorbent material, such as color changes, hardening, or fragmentation, are also signals for replacement. The replacement time provided by manufacturers is usually recommended based on standard usage conditions, and the actual service life may vary depending on frequency of use, operating environment, and maintenance.