A When adding materials to the automatic large transition compartment of the
Mikrouna glove box, click the "Automatic Material Addition Switch for Large Transition Cabin", and the system will automatically complete the process of adding materials. The specific process is as follows:
1) The outer door of the large transition compartment begins to open, and the buzzer starts to sound, reminding the operator to pay attention to safety;
2) After the outer door is opened, the placement of added items must be completed within the set delayed closing time;
3) After placing the three items, the system will delay the closing time and automatically start closing the outer door of the large transition compartment;
4) After the outer door of the large transition compartment is closed, the system automatically fills the large transition compartment according to the set parameters;
5) After the filling is completed, the inner door of the large transition compartment begins to open, and the required items need to be taken out within the set delayed closing time;
6) After the item is taken out, the system will delay the closing time and automatically start closing the inner door of the large transition compartment;
7) After the inner door of the large transition compartment is closed, if no operation is carried out within 10 minutes, the system will automatically draw the large transition compartment to a negative pressure of about 500mbar and place it.